Troubleshooting
5
Troubleshooting
5.1
Warnings
Fault
Cause
The residual run-
time of the
A segment of the
rechargeable bat-
battery status indi-
tery is low
cator is flashing
(< 30 minutes).
yellow.
The particle filter
residual capacity is
A segment of the
low (< 20 %).
particle filter resid-
ual capacity indica-
tor is flashing
yellow.
Malfunction during
switch-on (e.g.
A segment of the
caused by missing
flow indicator is
hose or filter).
flashing yellow.
5.2
Alarms
5.2.1 Blower unit
Fault
Cause
The residual run-
time of the
A segment of the
rechargeable bat-
battery status indi-
tery is almost
cator is flashing
exhausted
red.
(< 10 minutes)
The particle filter
residual capacity is
A segment of the
almost exhausted
particle filter resid-
(< 10 %)
ual capacity indica-
tor is flashing red.
Faulty breathing air
supply during oper-
A segment of the
ation (e.g. caused
flow rate indicator
by missing hose,
is flashing red.
filter or kinked
hose).
General system
error
One segment at a
time is flashing red.
10
5.2.2 Standard battery charger
Fault
Remedy
Status LED is
Recharge the bat-
flashing red.
tery soon or
replace with fully
charged battery.
Change particle or
combination filter
soon.
6
6.1
Re-check the
device function
Work to do
and prepare for
use.
Clean and disinfect the device
Carry out a visual inspection (see "Visual
Remedy
inspection", page 11)
Recharge the bat-
Carry out a tightness test (see Technical
tery or replace with
Manual)
fully charged bat-
tery.
Carry out a functional test (see "Check-
ing flow and warning devices", page 12)
Replace the O-ring at plug-in or bayonet-
Change particle or
type hose connector (see "Replacing the
combination filter.
O-ring at plug-in or bayonet-type hose
connector", page 12)
1) For gas-tight packed devices, otherwise every 6 months
2) Not stipulated by Dräger. Observe national guidelines where applica-
Re-check the
device function
and prepare for
use.
Device must be
checked by Dräger
Service.
Cause
General error or
defect
Maintenance
Maintenance intervals
2)
ble.
Instructions for use
Remedy
Re-insert the
rechargeable bat-
tery in the battery
charger. If the error
occurs repeatedly
have Dräger Ser-
vice check the bat-
tery charger and
rechargeable bat-
tery.
1)
X
1)
X
1)
X
1)
X
X
®
|
Dräger X-plore
8000
loading