Bioventus EXOGEN Guia Del Usuario página 20

Ocultar thumbs Ver también para EXOGEN:
Idiomas disponibles

Idiomas disponibles

EXOGEN Care
EXOGEN should be handled with care. Please
note the following:
• Use only a clean soft cloth, paper towel, or cotton
swab to clean EXOGEN, the transducer and the
strap. Do not use cleaning agents or solvents on
any of the components of the system
• Do not attempt to modify, disassemble or repair
the EXOGEN. There are no user serviceable parts
inside EXOGEN
• Exercise care when handling the transducer as
rough handling may scratch the transducer face
and cause EXOGEN not to work properly
• If any parts of EXOGEN or its accessories are
damaged, do not use EXOGEN. Please contact
Customer Service at 1-800-836-4080 to return
your EXOGEN for servicing
• EXOGEN is classified as an IP-22 device. The
IP-22 classification indicates that EXOGEN
provides:
- Protection against the access of fingers or
similar objects from the internal components
of EXOGEN
- Protection against the harmful ingress of
water into the enclosure of EXOGEN when
tilted up to 15° from normal position
• The EXOGEN transducer is classified as an IP-67
component. The IP-67 classification indicates that
the transducer is:
- Dust-tight
- Will not be damaged by water under defined
conditions of pressure and time (up to 1 meter
underwater)
• Never put EXOGEN in or under water
17
Operating Conditions
EXOGEN should be operated within:
Ambient temperature range: 5°C to 32°C (41°F
to 89°F)
Relative humidity range: 15% to 75% (non-
condensing)
Atmospheric pressure range: 700 hPA to
1060 hPA
Interference with proper operation of EXOGEN
may occur in the vicinity of equipment such as
portable and mobile communication units marked
with this symbol
. If abnormal operation of
EXOGEN is observed, attempt to relocate or
reorient EXOGEN in relation to the interfering
equipment until the interference stops.
The charger will function with an input voltage
range from 100 VAC -240 VAC and has an
operating frequency range of 50/60 Hz. The
charger output is 5 VDC.
EXOGEN and accessories should be stored
and transported within:
Ambient temperature range: 0°C to 32°C (32°F
to 89°F)
Relative humidity range: 15% to 85%
Atmospheric pressure range: 700 hPA
to 1060 hPA
If EXOGEN is stored or transported in
temperatures outside this range, allow
EXOGEN time to come to room temperature
for at least 30 minutes before operating.
The least favorable working conditions
for EXOGEN are +32
C at 75% Humidity.
o
Storage
• To prevent damage to EXOGEN and its
accessories, store EXOGEN in its carrying
case while not in use
• Do not store EXOGEN near radiators or
extreme heat
• Do not expose EXOGEN to extreme temperatures
or the internal electronic components may be
damaged
• As with any home electronic device, protect
EXOGEN from impact, exposure to moisture,
liquid spills, sand, dirt or debris
After your fracture has healed or prior to long-
term storage of EXOGEN, remove the battery to
prevent leakage of the battery.
EXOGEN Expected Service Life
The expected service life of EXOGEN and its
accessories is 343 treatments (6860 minutes).
Once EXOGEN delivers 343 treatments, it will
not provide further treatment.