HoMedics BPA-102DDM Manual De Instrucciones página 16

Idiomas disponibles

Idiomas disponibles

TROUBLESHOOTING
If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
Unit does not turn on when
Batteries have
the START/STOP
run down.
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on display or
incorrectly.
the blood pressure value is
displayed excessively low
Did you talk or move
(or high).
during measurement?
Shaking of the arm with
the cuff on.
Er error symbol shown on
Did you move during
display - measurement result
measurement?
exceeds measurement range.
Shaking of the arm with
the cuff on.
E1 error symbol shown on
Air circuit abnormality. Cuff
display
tube may not be plugged into
monitor correctly.
E2 error symbol shown on
Inflation pressure exceeding
display
300 mmHg.
E3 error symbol shown on
Data error.
display
– – appears on display
Recalibration alert.
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
CORRECTION
CAUSES
Replace them with
four new AA
alkaline batteries.
Re-insert the batteries in the
correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep arm
steady during measurement.
Measure again. Do not move
during measurement.
Check cuff connection.
Measure again.
Switch the unit off, then
measure again.
Remove and reload the
batteries.
Refer to page 25 for more
information.
29
SPECIFICATIONS
Four 1.5V DC (AA) batteries or
Power Source:
optional 12V DC AC adapter (not included)
Measurement Method:
Oscillometric
Pressure: 40~250 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories for each User (120 total)
Auto-shut-off:
1 minute after last button operation
Temperature: 50°F~104°F (10°C ~40°C)
Operation Environment:
Humidity: 40~85% RH max.
Temperature: 14°F~140°F (-10°C~60°C)
Storage Environment:
Humidity: 10~90% RH max.
Weight:
0.65 lbs (without batteries)
Arm Circumference
8.7"- 18.1" (22- 46cm)
:
4.3"(L) x 6"(W) x 2.9"(H)
Dimensions:
109mm(L) x 153mm(W) x 74 mm(H)
Type BF
Device and cuff are designed to provide special
protection against electrical shocks.
Note: These specifications are subject to change without notice.
30
loading