Stryker SDC3 Manual Del Usuario página 86

Sistema de gestión de información hd, paquete powershare, versión 2.2
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 491
P P r r o o b b l l e e m m
Application has
encountered a
critical error and is
shutting down.
No live video on
preview
No live video on
the monitor
No live video in
PIP on preview
No live video PIP
on monitor
Print errors
Error saving more
than 4GB files to
the USB
Error reading
archived cases
from DVD/CD
EN
84
S S o o l l u u t t i i o o n n
Please reboot the system.
Make sure that the camera is on and check the cable connections.
Make sure that the proper video source is selected.
Check the cable connections
Make sure that the camera is powered on.
Make sure that the proper video source is selected.
Check the cable connections
Make sure that the camera is powered on.
Make sure that the proper video source is selected.
Make sure that you are in PIP mode.
Check the cable connections.
Make sure that the camera is powered on.
Make sure that the proper video source is selected.
Make sure that the monitor OSD PIP is enabled via the application.
Make sure there aren't any paper jams.
Make sure the printer has enough ink.
Make sure the printer door is closed.
Make sure the printer has paper.
Make sure the printer is powered on.
Make sure the printer cable is connected..
Make sure that the USB is formatted as NTFS. FAT 32 will not support
any file size bigger than 4GB.
Reboot the system if you have just written the case. Check the disc on a
different system.
P23566L
loading

Este manual también es adecuado para:

0240060100