Remove power to the pump, so that the engine can't start accidentally.
DISASSEMBLY
Remove the transparent protection cover through the 2 screws.
Position the impeller with the 2 white rollers in a vertical position.
Remove from its seat the ring on the left and then pull the pipe upward.
Manually rotate the rotor clockwise in order to pull up even the connection put at the right
of the pump.
ASSEMBLY
Position the impeller with the 2 white rollers in a horizontal position.
Press the pipe into his place, following gradually the path.
Remove from its seat the ring on the left and then pull the pipe upward.
Manually rotate the rotor clockwise in order to enter also the right connection of the pump.
Enter again the transparent protection cover, pressing it to its place and tighting the 2
screws.
8
• Carry out a visual check to verify the correct installation or any possible leak.
• Run at least a loop to test the success of the operation.
• If necessary execute an electrical test (EN 61010)
REPLACEMENT OF PUMP PIPE
TESTING PROCEDURE
At the end of any assistance service is a good rule: