P7
1. Remove the front panel and the power board following procedures P1 and P4.
2. Then extract the fittings from the pump, extract fastons from contacts and remove
the self-locking nuts.
3. Replace the drain pump and remount it by fixing it on the bracket at first and then
carrying out the electrical and hydraulic connections.
Replacement of the drain PUMP
RP
ENGLISH