ERR 008
Remove the drain cap from the drain coupler
Check that there are no breaking or obstructions in the drain pipe. Replace pipe or
remove obstructions
Clean the filter inner the tank.
Check electrical continuity and integrity of the cable connecting EV drain and Main Board.
If necessary, replace the cable.
Replace the board.
ERR 009
It occurs if the PT100 probe does not perceive an increase of 1°C in 1 minute
Replace the PT100
Replace the board
Check operation of the fan and if necessary replace it
Check operation of the armored resistance and if necessary replace it
PROBLEM DURING WATER DRAIN PHASE
It occurs if there are the following conditions:
If at the end of the drain the tank signals MAX level
HOW TO OPERATE
PROBLEM DURING DRYING PHASE
HOW TO OPERATE