P8
Before replacing the exhaust EV control the ohmic value.
1. Remove the front panel and the power board following procedures P1 and P4.
2. Detach electric connections and the Teflon hose from the solenoid valve (the hose
is attached through a John Guest quick coupler).
3. Dismount the solenoid valve by intervening on the internal nut and on the external
screw.
4. Remount the new solenoid valve by following the above procedure in the opposite
sequence.
38
Disassembly of the charge SV
NOTE
Set the meter value 20k -> Ω
The correct value is 3.95 on the display
loading

Este manual también es adecuado para:

0zvmt0001