3.4 Instrument Setup

The instrument allows the user to access and modify the following
parameters through direct button interface:
Calibration Options
Alarm Options
Instrument Options.
These menus can be accessed only from the MEASURE page by
pressing and holding the ▲ and ▼ buttons simultaneously until
prompted for a password.
The operation is as follows:
1. Turn the instrument ON and wait until the MEASURE page
appears.
2. Simultaneously press and hold the ▲ and ▼ buttons for
approximately five seconds.
• The default password is "672".
3. Enter the first digit by pressing the ▲ or ▼ button, and confirm
with the
• The cursor jumps to the second digit.
4. Enter the second and third digits.
• Incorrect password: Instrument returns to the MAIN page.
• Correct password:
The password can be changed with a PC through the
MSA Link software.
If the password is forgotten, it can be reset by using MSA Link software.
Contact MSA Customer Service for assistance.
The following Options are available by pressing the ▲ and ▼ buttons:
Calibration Options - see Section 3.4.1
Alarm Options - see Section 3.4.2
Instrument Options - see Section 3.4.3.
3-6
Figure 3-2. Password Screen
button.
User can enter the Setup mode.