3.9 Calibration with the Galaxy Test System
The instrument can be calibrated using the Galaxy Automated Test
System - contact MSA for a list of compatible gases and concentrations.
NOTE: The Galaxy is not a CSA certified method of calibration.
Similar to the successful (manual) calibration described in Section 3.8.3,
a √ symbol displays on the MEASURE page after successful
Galaxy calibration.
•
This √ symbol:
•
appears on the color display in the upper feature bar
•
appears on the monochrome display in the lower right corner
•
remains on the display for 24 hours after the calibration
•
goes OFF after 24 hours.
NOTE: If the horn alarm is turned OFF, the calibration √ symbol
does not appear on the color display.
3.10 Instrument Shutdown
For instrument shutdown, press and hold the
•
The instrument:
•
displays a flashing "HOLD BUTTON FOR SHUTDOWN"
•
shows a progress bar, informing the user how much longer
to hold the button to complete the shutdown.
3-24
Figure 3-9. Instrument Shutdown
button.