3.9 Calibration with the Galaxy Test System

The instrument can be calibrated using the Galaxy Automated Test
System - contact MSA for a list of compatible gases and concentrations.
NOTE: The Galaxy is not a CSA certified method of calibration.
Similar to the successful (manual) calibration described in Section 3.8.3,
a √ symbol displays on the MEASURE page after successful
Galaxy calibration.
This √ symbol:
appears on the color display in the upper feature bar
appears on the monochrome display in the lower right corner
remains on the display for 24 hours after the calibration
goes OFF after 24 hours.
NOTE: If the horn alarm is turned OFF, the calibration √ symbol
does not appear on the color display.

3.10 Instrument Shutdown

For instrument shutdown, press and hold the
The instrument:
displays a flashing "HOLD BUTTON FOR SHUTDOWN"
shows a progress bar, informing the user how much longer
to hold the button to complete the shutdown.
3-24
Figure 3-9. Instrument Shutdown
button.