Cleaning: Automated
1. Brush
•
Using a syringe, inject 50mL of the enzymatic detergent (from the
"Preparation for Cleaning" section) into any lumen and mated surface
a minimum of one time.
•
Brush from both ends of any lumens a minimum of 1 time, using an
appropriate bottle brush.
2. Rinse
•
Rinse the device with RO/DI water at ambient temperature until there
is no visible detergent residue. Continue to rinse for a minimum of
30 seconds after all detergent residue has been removed.
•
Place the device in the washer on an incline to facilitate drainage.
3. Automated wash
•
Program the washer using the following parameters:
Phase
Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry Phase
•
Upon completion of the draining phase after Rinse 1, stop the cycle and
open the washer door.
•
Remove the device from the washer during the thermal phase and place
the device back into the washer for the dry phase.
•
If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
ENZOL® Enzymatic Detergent is validated for cleaning efficacy.
2
Renu-Klenz® is validated for cleaning efficacy.
3
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Water
Temperature
Cold tap
water
Hot tap water
Set point
(66°C)
Hot tap water
115°C
Detergent Type and
Concentration
(if applicable)
N/A
Enzymatic Detergent
Non-enzymatic Detergent
N/A
N/A
2
3
17