Cleaning: Automated
1. Brush
Using a syringe, inject 50mL of the enzymatic detergent (from the
"Preparation for Cleaning" section) into any lumen and mated surface
a minimum of one time.
Brush from both ends of any lumens a minimum of 1 time, using an
appropriate bottle brush.
2. Rinse
Rinse the device with RO/DI water at ambient temperature until there
is no visible detergent residue. Continue to rinse for a minimum of
30 seconds after all detergent residue has been removed.
Place the device in the washer on an incline to facilitate drainage.
3. Automated wash
Program the washer using the following parameters:
Phase
Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry Phase
Upon completion of the draining phase after Rinse 1, stop the cycle and
open the washer door.
Remove the device from the washer during the thermal phase and place
the device back into the washer for the dry phase.
If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
ENZOL® Enzymatic Detergent is validated for cleaning efficacy.
2
Renu-Klenz® is validated for cleaning efficacy.
3
Recirculation
Time
2 minutes
2 minutes
2 minutes
2 minutes
7 minutes
Water
Temperature
Cold tap
water
Hot tap water
Set point
(66°C)
Hot tap water
115°C
Detergent Type and
Concentration
(if applicable)
N/A
Enzymatic Detergent
Non-enzymatic Detergent
N/A
N/A
2
3
17