Keysight U1211A Guía Del Usuario Y Servicios página 91

Multímetros de pinza
Ocultar thumbs Ver también para U1211A:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Table 6-2
Performance verification tests (continued)
Test function
AC voltage
Peak voltage (max)
Frequency
Duty cycle
[c]
AC + DC voltage
Keysight U1211A, U1212A, and U1213A User's and Service Guide
Range
Reference signals/
5520A output
4 V
4 V, 45 Hz
4 V, 2 kHz
40 V
40 V, 45 Hz
40 V, 2 kHz
400 V
400 V, 45 Hz
400 V, 400 Hz
400 V, 2 kHz
1000 V
1000 V, 45 Hz
1000 V, 400 Hz
1000 V, 2 kHz
400 V
400 V
99.99 Hz
10 Hz, 0.6 V
9.999 kHz
2 kHz, 20 V
0.1% to 99.9%
5 V
square wave, 2 kHz
4 V
4 V, 45 Hz
4 V, 2 kHz
40 V
40 V, 45 Hz
40 V, 2 kHz
400 V
400 V, 45 Hz
400 V, 2 kHz
1000 V
1000 V, 45 Hz
1000 V, 2 kHz
Performance Tests and Calibration
Error limits (from nominal 1 year)
values
U1211A
±4.5 V
±4.5 V
±15 V
±15 V
, 60 Hz
±8.3 V
p
±0.007 kHz
at 50%,
pp
U1212A
U1213A
±0.045 V
±0.085 V
±0.45 V
±0.85 V
±4.5 V
±4.5 V
±4.5 V
±8.5 V
±15 V
±15 V
±15 V
±25 V
±8.3 V
±8.3 V
±0.05 Hz
±0.007 kHz
±0.9%
±0.069 V
±0.109 V
±0.69 V
±1.09 V
±6.9 V
±10.9 V
±24 V
±34 V