Table 6-6
U1212A valid adjustment reference input values (continued)
Function
Range
Temperature
Short
0.4 V
K-type
Diode
Short
2.000 V
Table 6-7
U1213A valid adjustment reference input values
Function
Range
DC voltage
Short
4 V
40 V
400 V
1000 V
AC voltage
4 V
40 V
400 V
1000 V
Keysight U1211A, U1212A, and U1213A User's and Service Guide
Reference input value
SHORT
0.400 V
000.0 °C
SHORT
2.000 V
Reference input value
SHORT
3.000 V
30.00 V
300.0 V
1000 V
0.200 V (70 Hz)
3.000 V (70 Hz)
3.000 V (2 kHz)
030.0 V (70 Hz)
30.00 V (70 Hz)
30.00 V (2 kHz)
030.0 V (70 Hz)
300.0 V (70 Hz)
300.0 V (2 kHz)
100 V (70 Hz)
1000 V (70 Hz)
1000 V (2 kHz)
Performance Tests and Calibration
Valid range for reference input
Short V and COM terminals
0.9 to 1.1 × reference input value
Provide 0 °C with ambient compensation
0 Ω
0.9 to 1.1 × reference input value
Valid range for reference input
Short V and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
6
103