Selecting the adjustment mode
To unsecure the instrument, see
page 97 or
unsecured, the reference value will be indicated on the primary display.
Entering adjustment values
Use the following adjustment procedure to enter an input calibration value from
the front panel:
1 Press
2 Press
3 Press Hold/Max Min when you are done.
Verify the adjustments using
Table 6-10
Table 6-8
U1211A list of adjustment items
Function
AC voltage
DC voltage
Keysight U1211A, U1212A, and U1213A User's and Service Guide
"Resetting the security code to factory default"
or
(refer to
primary display.
or
(refer to
to 9.
for U1213A.
Range
400 V
1000 V
SHrt
400 V
1000 V
Performance Tests and Calibration
"Unsecuring the instrument for calibration"
Table 4-1
on page 64) to select each digit in the
Table 4-1
on page 64) to advance through the digits 0
Table 6-8
for U1211A,
Ad justment item
30.00 V (70 Hz)
300.00 V (70 Hz)
300.00 V (2 kHz)
100.0 V (70 Hz)
1000.0 V (70 Hz)
1000.0 V (2 kHz)
Short
300.0 V
1000 V
on page 99. Once
Table 6-9
for U1212A, and
6
on
109