6
Performance Tests and Calibration
Table 6-2
Performance verification tests
Test function
[a]
Temperature
Resistance
Diode
Capacitance
DC voltage
90
Range
Reference signals/
5520A output
–200 °C to –40 °C
–40 °C to 1372 °C
–40 °C to 1372 °C
400 Ω
4 kΩ
40 kΩ
400 kΩ
4 MΩ
40 MΩ
Diode
4 µF
40 µF
400 µF
4000 µF
4 V
40 V
400 V
1000 V
Keysight U1211A, U1212A, and U1213A User's and Service Guide
Error limits (from nominal 1 year)
values
U1211A
–200 °C
–
0 °C
–
1372 °C
–
400 Ω
±2.3 Ω
4 kΩ
±0.023 kΩ
40 kΩ
–
400 kΩ
–
4 MΩ
–
40 MΩ
–
1.9 V
±0.012 V
4 µF
–
40 µF
–
400 µF
±8.4 µF
4000 µF
±124 µF
4 V
–
40 V
–
400 V
±2.5 V
1000 V
±8 V
U1212A
U1213A
± 5.0 °C
± 5.0 °C
± 1.0 °C
± 1.0 °C
± 14.7°C
±14.7°C
±2.3 Ω
±1.5 Ω
±0.023 kΩ
±0.015 kΩ
–
±0.15 kΩ
–
±1.5 kΩ
–
±0.027 MΩ
–
±0.85 MΩ
±0.012 V
±0.012 V
–
±0.044 µF
–
±0.44 µF
±8.4 µF
±8.4 µF
±124 µF
±124 µF
–
±0.013 V
–
±0.13 V
±2.3 V
±1.3 V
±8 V
±8 V