6
Performance Tests and Calibration
Table 6-6
U1212A valid adjustment reference input values
Function
Range
DC voltage
Short
400 V
1000 V
AC voltage
400 V
1000 V
DC current
Open
40 A
400 A
1000 A
AC current
40 A
400 A
1000 A
Resistance
Short
4 kΩ
400 Ω
Capacitance
400 µF
4000 µF
102
Reference input value
SHORT
300.0 V
1000 V
030.0 V (70 Hz)
300.0 V (70 Hz)
300.0 V (2 kHz)
100 V (70 Hz)
1000 V (70 Hz)
1000 V (2 kHz)
OPEN
30 A
300 A
300 A
02.00 A (70 Hz)
30.00 A (70 Hz)
030.0 A (70 Hz)
300.0 A (70 Hz)
50 A (70 Hz)
300 A (70 Hz)
SHORT
3.000 kΩ
300.0 Ω
300.0 µF
3000 µF
Keysight U1211A, U1212A, and U1213A User's and Service Guide
Valid range for reference input
Short V and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Keep jaw closed without conductor
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
Short Ω and COM terminals
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value
0.9 to 1.1 × reference input value