5. Commissioning, operation

5.4 Permissible ambient and operating temperatures

When mounting the pressure gauge it must be ensured that, taking
into consideration the influence of convection and heat radiation,
EN
no deviation above or below the permissible ambient and media
temperatures can occur. The influence of temperature on the indica-
tion accuracy must be observed.
In order to avoid any additional heating, the instruments must not be
exposed to direct solar irradiation while in operation.
5.5 Permissible vibration load at the installation site
The instruments should always be installed in locations free from
vibration.
5.6 Commissioning
During the commissioning process pressure surges must be avoided
at all costs.
5.7 External zero point setting (only model PG43SA-S with
vent hole)
Setting range ±15° = ±5.5 % of the span
External zero point setting is not available for the model PG43SA-S
with vent valve.
14
WIKA operating instructions pressure gauge, PG43SA-S, PG43SA-C
The zero point setting can be
made with a slotted screwdriver
after the removal of the sealing
plug on the top of the case.