Cleaning And Disinfection; Maintenance And Repair - Olympus CelonLab ENT Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
Instructions for Use
4. Instructions for Use
Caution: The power control unit is intended for use by qualified medical
personnel trained in the use of electrosurgical equipment and targeted
puncture of areas of organs. Please also observe Section 1 (Safety and
Warnings) and also the Instructions for Use for the coagulation electrodes and
accessories.
1. Connect the foot switch to the socket (3).
2. Switch on the control unit at the power switch (1). The unit is ready for
operation after approx. 10 seconds.
3. Connect an RFITT coagulation electrode to the socket (2).
4. Set the necessary coagulation power with buttons (7) and (8) in the range
between 1 and 25. The power depends on the type of treatment in
question. Suggestions for suitable settings are to be found in the
Instructions for Use of the compatible coagulation electrodes.
5. Insert the coagulation electrode into the tissue area to be treated.
6. Activate the power output by pressing the footswitch (the footswitch must
be held down throughout the entire treatment).
7. End the application by releasing the footswitch when the end of treatment
(dehydration of the tissue) is indicated by a pulsed acoustic signal or at
any time during the treatment.
8. Repeat steps 4 to 7 if required by the therapy.
9. Switch off the unit at the power switch (1).
10. Clean the components following the instructions in Section 5.
Caution: When using bipolar radiofrequency induced thermotherapy (RFITT),
a lower power level may lead to longer application times and therefore a larger
coagulation volume.

5. Cleaning and Disinfection

All surfaces of the unit's housing can be cleaned and disinfected with the
cleaning agents and surface disinfectants normally used for medical equipment
(mild cleaning solution, e.g. 70 % isopropyl alcohol). No liquid must enter the
connector or the unit during cleaning.
The CelonFoot
mentioned above.

6. Maintenance and Repair

6.1 Changing the Power Fuses
The unit is protected by two power fuses accessible from the outside. If these
fuses blow, an authorised technician should check the unit for possible defects
before new fuses are fitted and the unit is operated again, as overloading of the
fuses is usually the result of a defect.
Caution: Disconnect the power plug before changing the fuses! The fuses
inside the housing must only be changed by an authorised technician.
W7098046
EN
Switch
can be spray or wipe disinfected with the products
ENT
CelonLab
9
Tabla de contenido
loading

Tabla de contenido