Olympus CelonLab ENT Instrucciones De Uso página 7

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
Instructions for Use
1
Power switch: The power switch is for switching the unit on and off.
When the power control unit is switched on the power switch glows green.
For safety reasons the unit is only ready for use after approx. 10 seconds.
The power setting is always at its minimal value of 1 after switching on.
2
Connection sockets for RFITT coagulation electrodes: All compatible
bipolar coagulation electrodes can be connected to this socket.
3
Connection socket for the foot switch: The CelonFoot
connected to this socket. When the footswitch is pressed, power output is
activated. The generator cannot be operated without a footswitch
connected.
4
Signal lamp: "Power output": If the control unit is activated by pressing
the footswitch, and if power can be emitted at the output, this is indicated
by this blue lamp.
5
Signal lamp: "resistance too low": This red lamp illuminates, in
combination with an acoustic signal (intermittent tone), to indicate that the
resistance is lower than 30 ohms. Possible cause: short circuit in the
electrode or cable. For safety reasons the power output is automatically
switched off.
6
Power display: This two-digit LED display indicates the chosen power
setting (values between 1 and 25). The indicated value does not necessarily
correspond to the emitted power in watts because the power depends on
the tissue resistance.
7
Button for increasing the power setting: Pressing this button increases
the power level by 1.
8
Button for reducing the power setting: Pressing this button decreases
the power level by 1.
9
Symbol: "Unit is defibrillator-safe": The use of a defibrillator will not
cause any damage to the unit, provided that the power control unit is not
active, i.e. the footswitch is not pressed.
10 Symbol: "Caution! Read the Instructions for Use": If a unit carries
this symbol, it is necessary to read the Instructions for Use before using it
for the first time.
11 Connection for equipotential bonding: See Chap. 3.2 (Installation
Instructions)
12 Power connection: This socket is used for connecting the unit to the
power supply system. (see Chap. 3.1)
13 Holder for power fuses: The power control unit is protected by two
external fuses (see Chap. 6.1). If these fuses blow, an authorised technician
should check the unit for possible defects before new fuses are fitted and
the unit is operated again.
14 Rating plate: Details of the manufacturer, supply voltage, maximum
power input, moisture protection class (IP), serial number and relevant
output data.
15 Volume switch: This switch is used for setting the volume of the acoustic
signal to loud or quiet. In the standard setting (switch not pressed) the
signal is loud. The sound can be decreased by pressing the switch.
16 Symbol: "Non-ionising radiation": Units which carry this symbol emit
radiofrequency energy for diagnostic or therapeutic purposes.
W7098046
EN
ENT
CelonLab
Switch
can be
7