Perform sensor calibration via HART handheld terminal
14
Select Sensor Calibration menu.
1.
2.
Select calibration method.
3.
For O2 calibration select calibration unit.
4.
The following message is shown "Capture current measured value".
5.
Press [OK] to store the current measured value.
6.
The following message appears "Captured value is stored. Take a grab
sample to measure in the lab or perform parallel measurement." Press [OK].
7.
The following message is shown "Enter reference value". You can perform this
step at any time. Press [OK].
8.
The captured value is shown.
9.
Enter the measured reference value.
10. Press [OK] to store the reference value.
11. If the reference value is within the valid range, the Slope and Offset are
shown. Press [OK].
12. Select Adjust, Calibrate or Abort. Press [OK].