Perform sensor calibration via configuration tool or asset-management tool
For calibrating the sensor in the lab or in non-hazardous areas you can use iSense.
The scope of delivery comprises the iSense software on CD-ROM and the iSense
cable.
1.
Select Sensor Calibration menu.
2.
Select calibration method.
3.
For O
calibration select calibration unit.
2
4.
Click [Step 1: Capture current measured value].
5.
The current Sensor value and Status are shown.
Click [Next] to store the measured value.
6.
7.
The following message appears "Captured value is stored. Take a grab
sample to measure in the lab or perform parallel measurement." Click [OK].
8.
Click [Step 2: Enter reference value]. You can perform Step 2 at any time.
9.
The captured value of "Step 1" is shown.
10. Enter the measured reference value.
11. Click [Next] to store the reference value.
12. If the reference value is within the valid range, the Slope and Offset are
shown. Click [OK].
13. The following message appears "Complete calibration procedure.
Select either "Adjust", "Calibrate" or "Abort".
14. Select Adjust, Calibrate or Abort. Click [OK].
13