HART handheld terminal
h
9.
Calibrate sensor. Menu path: Device Setup > Sensor Calibration
10. Perform further settings. See manual of the M100 transmitter.
11. Store configuration to device.
Note: The DD "008E8E7D0101.hhd" is on the supplied CD-ROM. You can also
download the DD via Internet "www.mt.com/M100".
For step 1 refer to the documentation about the HART handheld terminal.
1.
Check if the DD of the M100 transmitter has already been installed on the
HART handheld terminal. Install the DD if necessary.
2.
The communication is built up automatically.
3.
Load configuration from device. Menu path: Device Setup > Detailed Setup
Set Tag or/and Long tag.
4.
Menu path: Device Setup > Detailed Setup > HART Info
Set Date and Time. Set the time in the 24 hour format. The time format can-
5.
not be changed. Menu path: Device Setup > Detailed Setup > System
6.
Set range of the analog output signal.
Menu path: Detailed setup > Output Condition > Analog output > Range
URV (Upper range value) and LRV (Lower range value), can be changed
–
to the current measurement range
USV (Upper range sensor) and LSV (Lower range sensor), defined by the
–
sensor and cannot be changed
Define the process variables PV, SV, TV and QV
7.
Menu path: Device Setup > Detailed Setup > Measurements > Channel Setup
8.
Calibrate sensor. Menu path: Device Setup > Sensor Calibration
9.
Perform further settings. See manual of the M100 transmitter.
11