Oscilloscope Section/English
1. Store the current displayed waveform into the device.
•Press "HOLD"
"F2(Save)"
enter into "Save Operations" state
→
→
"F1(Memory)", when no store exists in the position(No store), store directly, the color may
→
turn to purple red from green if storage succeeds.
•If previous store exists, system presents "Into the new?" if answer "yes", original content is
covered, current content is stored; if "no", quit the operation.
•Press "F4 (EXIT)" to quit.
2.Store the record state waveform into the device.
•Press "REC"
"F2 (Save)"
"
"(select position)
→
→
as 1.
•Press "F4 (back)" to turn back to flip operation interface of record state, press "REC"to quit.
3.Store FFT displayed waveform into device.
•Press "F4 (FFT)"
"HOLD"
"F1 (Save)"
→
→
steps same as 1.
2 15 2 Deletion
- -
Under "Save Operations" state, press "
system presents "Are you sure?" if answer "yes", the content is deleted; if "no", quit the
operation.
2-15-3 Callout
•Under "Save Operations" state, press "
the selected content displays on the screen.
•The waveform data is sent to upper computer by Bluetooth or wireless USB while waveform
callout.
Picture 2-13-1
44
"
" (select position)
→
"F1 (Memory)", rest steps same
→
"
" (select position) "F1 (Memory)", rest
→
""
"(select position), then press "F2(Delete)",
""
"(select position), then press "F3 (Recall)",
Picture 2-13-2
13/08/2017 Version No. 001
•Press F2 (return) under waveform displaying state, turn back to "Save Operations" state.
•Press "F4 (EXIT)" in waveform displaying state, turn back to waveform displaying state.
3.
Examples
This chapter mainly introduces several application examples, these simplified examples focus
on some main functions of the oscilloscope, and user may take it for reference to solve some
actual testing problems.
•Simple measurement
•Cursor measurement
•Analyze detailed information of signal
•Analyze communication signal difference by mathematical computing function
3 1 Simple Measurement
-
Observe unknown signal in circuit, fast display the frequency and peak value of the measured
signal.
1. Auto search function
In order to fast display the signal, take following steps:
(1) Connect correctly the probe and connector.
(2) Press "AUTO", wait a moment, the oscilloscope auto searches most of the displayed signals.
User can take manual adjustment on this basis, until the displayed waveform satisfies user's
needs.
Note: the oscilloscope displays corresponding auto search result in waveform area on the
screen according to detected signal type.
3-2 Cursor measurement
The oscilloscope can auto measure most displayed signals. If needs to accurately measure
the frequency and peak value of signal, take following steps:
(1)measure signal frequency
•Press "HOLD", freeze the waveform.
•Press " " or " ", to move left cursor.
•Press "MODE", to switch cursor, press " " or " ", to move right cursor.
•Read the time (cycle) between the two cursor press "MODE", read corresponding frequency.
(2) measure signal peak value
•Press "F1 (V/F cursor)" after step 1, switch to horizontal cursors, press " "or" " to move
up cursor.
•Press "MODE", to switch cursor, press " "or" " to move down cursor.
•Read the voltage value between two cursors.
13/08/2017 Version No. 001
Oscilloscope Section/English
45