Technical data

Batteries must not be disposed of as household waste.
They are therefore marked with the adjacent symbol.
Dispose of batteries at battery collection centres as
specified by the applicable regulations.
10
Technical data

10.1 Gas detector

Ambient conditions:
during operation and storage -20 to +50 °C
Degree of protection
Alarm volume
Orientation
Gas detector storage time
Sensor storage time
Flow direction of air in diffu-
sion operation
Battery
Dimensions
Weight (with pump)
Update interval for the dis-
play and signals
1) Tested without the pump or calibration adapter.
Operating time in normal conditions (diffusion mode)
With CatEx and 3 EC sen-
sors
With 3 EC sensors
42
700 to 1300 hPa (measuring
function)
800 to 1100 hPa (use in
explosion-hazard areas)
10 to 90 % (briefly up to
95 %) rel. hum.
1)
IP 68
Typically 100 dB (A) at a dis-
tance of 30 cm
Any
1 year
The ambient conditions and
the storage time for sensors
in their original packaging
correspond to those of the
gas detector
As part of the technical suit-
ability test: ≤6 m/s
Lithium-ion battery, recharge-
able, 3.6 V, 6.4 Ah, 24 Wh,
250 g
approx. 179 x 77 x 42 mm (H
x W x D)
Typically 550 g, depending
on which sensors are
equipped, without transport
harness
1 s
typically 24 h
typically 120 h
1) The nominal operating time of the gas detector at an ambient tem-
perature of 20 to 25 °C, 1013 mbar, alarms less than 1 % of the time,
power-save mode display activated. The actual operating time varies
depending on the ambient temperature and pressure and battery and
alarm conditions.
In pump mode: If the gas detector is used permanently in
pump mode, the operating time is reduced by approx. 2 h.