Tabla De Contenido; Specifications; General Information - Hach MET ONE 6000P Manual De Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 47
Table of Contents
Installation
on page 7
Startup
on page 13

Specifications

Specifications are subject to change without notice.
Specification
Details
Dimensions (W x D x
17.15 x 11.43 x 31.75 cm (6.75 x 4.5 x 12.5 in.)
H)
Enclosure
304 stainless steel
Light source
Long Life Laser
Weight
4.6 kg (10.2 lb)
Pollution degree
2
Installation category
DC models: I
AC models: II
Protection class
DC models: III
AC models: II
Power requirements
DC models: 24 VDC ± 10%, 60 W maximum
AC models: 100-240 VAC, 50-60 Hz, 0.4 A maximum
Operating temperature
5 to 40 °C (40 to 104 °F); best performance: 10 to
32 °C (50 to 90 °F)
Storage temperature
-40 to 70 °C (-40 to 158 °F)
Operation
on page 13
diode, Class 3B laser
Specification
Details
Humidity
Operating: 80% up to 31 °C (88 °F) decreasing linearly
to 50% relative humidity at 40 °C (104 °F)
Storage: 5 to 98% relative humidity, non-condensing
Altitude
2000 m (6562 ft) maximum
Fitting sizes
Model 6015P: Barb fitting for 1/4-in. ID inlet tubing, 3/8-
in. ID exhaust tubing; optional 3/8-in. ID inlet fitting
Output signal options
Pulse, analog 4-20 mA, serial RS232 with Modbus
RTU or FXB communication protocol (no networking),
serial RS485 with Modbus RTU or FXB communication
protocol, Ethernet with ModbusTCP protocol
Data storage
1000 samples (records), oldest records are written over
when the buffer is full
Sample flow rate
Model 6015P: 1.0 cfm (28.3 L/min) ± 5%
Range
Model 6015P: 0.5 μm to 10.0 μm at 1.0 cfm
(28.3 L/min)
Sensitivity
Model 6015P: 0.5 μm at 1.0 cfm (28.3 L/min)
Counting efficiency
Model 6015P: 50% ± 20% at 0.5 μm and 100% ± 10%
for sizes at least 1.5 times, or larger than the smallest
channel size
Coincidence loss
Model 6015P: 10% at 20,000,000 particles/m
(566,000 particles/ft
False count rate
One or less in 5 minutes
Pump type
Air vacuum, rated for continuous use
Certifications
Certified to UL and CSA by ETL, CE
1
Fully complies to JIS and ISO 21501-4.

General information

In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
1
.
3
3
)
English 3
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Met one 6015p