13 Guidelines for IQ and OQ
Some laboratories need the installer make an IQ document to certificate the good
installation of this equipment. As well as an OQ document to certificate the good
working of this equipment.
The installer can elaborate those documents following the examples:
Checking list for IQ (Installation Qualification):
•
Installer name.
•
Manufacturer name:
•
Unit identification: Code, model and serial number.
•
Unit location: Laboratory, Person in charge.
•
Checking packing list:
•
Unit location matches the requirements?
Water and electricity requirements are suitable?
•
Electrical supply.
Socket:
- Ground terminal?
- Suitable to power consumption. (16A)
- Voltage according plate ±10%
•
Water connection.
- Water pressure bigger than 1bar.
- Hose to waste are properly fixed and not blocked.
•
Water from steam generator: Low level of calcium, conductivity above 20
MicroSiemens.
•
Personnel training: Operator personnel understand the instruction manual?.
•
Date and name of who has made the Installation (Signed).
Checking list for OQ (Operation Qualification):
•
Installer name.
•
Manufacturer name:
•
Unit identification: Code, model and serial number.
•
Unit location: Laboratory, Person in charge.
•
Check the equipment distillate correctly. See: 8.1
•
Check nitrogen recuperation with ammonia sulphate sample. See 8.2.
Note : ammonia sulphate weight, HCl normality, HCl consumption and obtained
recovery.
•
Check the white test See: 8.3
Note: Ammonia sulphate weight, HCl normality, HCl consumption and obtained
recovery.
•
You can check the nitrogen recovery with acetanilide sample. See: 8.4.
Note: Acetanilide weight, HCl normality, HCl consumption and obtained recovery.
•
Date and name of who has made the Installation (Signed).
14 Warranty
The warranty covers the defective components or faulty assembling of the unit for
1 year since the purchasing date.
Keep purchasing bill as the only valid document for the purchasing date.
The warranty does not cover the damages caused for a unit malfunction. The unit
must be suitably qualified before to be used for valuable samples processing.
Return good are not accepted without previous J.P. SELECTA, s.a.u. approval.