Status Indication; Manual Operation - Selecta PRO-NITRO S Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CODIGO 80188 REV E
1
2
3
4
6
7
8
Fig. 5.1 Control panel

Status indication

1. Heater ON
2. Water feed fail.
3. Safety door open or sample tube
not present.
4. Overtemp on steam generator.
Display and selection keys.
5. Display.
5. Up/down keys.
Function keys.
6. Operation function mode selection
between «Manual» and «Automa-
tic».
7. H3BO3 / START
«MANUAL» mode: Boric dose.
«AUTO» mode: cycle start.
8. NaOH
Modo «MANUAL»: NaOH dose.
Modo «AUTO»: No function.
9. STEAM
«MANUAL» mode: Distillation.
«AUTO» mode: No function.
10. EMPTY/STOP
«MANUAL» mode: Empty tube.
«AUTO» mode: STOP
J
.P. SELECTA s.a.u.
Setting the dosing volume of NaOH:
5
- Press
- Keep pressed keys
- Press
- Use
(factory setting 12).
9
10
- Use
- Press
- Approximately each second means 5ml.
Setting the dosing volume of Boric acid:
- Press
- Keep pressed keys
- Press
- Use
(factory setting 4).
- Use
- Press
- Approximately each second means 10ml.

5.4 MANUAL operation

Select MANUAL mode. Key
Set parameters according 5.3.
Insert a sample tube (See 5.1) and an Erlenmeyer flask.
Adding Boric: Press
Boric acid will be added on the Erlenmeyer flask.
Adding NaOH: Press
NaOH will be added on the sample tube.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
Jun/2018
(Sujetas a modificaciones sin previo aviso)
.
to select NaOH.
to set the NaOH pump working time
to validate the setting.
to save the setting and exit from settings mode.
.
to select H3BO3.
to set to set the Boric pump working time
to validate the setting.
to save the setting and exit from the settings mode.
toggles working mode. See Fig 5.1
for a 10s until display shows dashes.
for a 10s until display shows dashes.
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 39