INSTRUCTION MANUAL CODE 80315 REV B
FREE SPACE AROUND CENTRIFUGE
Safety zone:
Leave a free space of 30 cm. around
the centrifuge.
With the centrifuge running:
• Do not place hazardous substances
in the security area.
• The staff should be outside the safe-
ty area.
J
.P. SELECTA s.a.u.
October 2019
5 Introduction
The centrifuge has an induction motor maintenance free and a maximum noise level
of 60 dBA. It has an internal security camera and a bucket, stainless steel cover.
The electronic circuit is controlled by a microprocessor that controls the operating
parameters: speed, acceleration, braking, RCF, temperature and operating time. It
can store up to 10 operating programs.
6 Installation
6.1
Location
•
Place the equipment near a power socket suitable for consumption and over a flat
level surface and strong enough to support the equipment weight and possible
vibrations.
•
The surface must allow the adherence of the centrifuge legs.
•
Leave a free space of 30 cm. around it (See danger indication).
•
This equipment can cause vibrations that could affect other nearby equipments.
•
Do not install the centrifuge near heat sources or in places where it is exposed
to direct sunlight, as this may affect temperature regulation.
•
Place the nipple provided in outlet (16) and connect to a hose to drain conden-
sates.
DO NOT INSTALL THE EQUIPMENT IN EXPLOSIVE ATMOSPHERES OR PLA-
CES WHERE SUBSTANCES LIABLE TO FORM EXPLOSIVE OR FLAMMABLE
MISXTURES ARE USED OR STORED.
6.2
First lid opening and emergency opening
Upon receiving the centrifuge, the network connection cable is inside the tank and
therefore the only way to open the equipment is following these instructions:
-
Remove the equipment side plastic caps.
-
Gently push from both caps and lid will open.
USE THIS OPENING SYSTEM FOR EMERGENCY OPENING ONLY.
NEVER use the device emergency opening when the
engine is rotating.