TM 530
Produzido por Tecnomotor Eletrônica do Brasil S/A - REPRODUÇÃO PROIBIDA - Eventuais erros ou defeitos favor comunicar
[email protected]
Remarks:
The purpose of this function is to facilitate the visualization and
identification of unknown signals. It may take a few tries until you get a
good signal analysis, especially in situations with an incidence of noise and
electromagnetic interference.
Where the identification (??) Appear, or if there is doubt about viewing
signal, try to activate the key
again for a new signal analysis. A more
reliable can be achieved in ways (stp) and (con), manually setting up the
parameters of the oscilloscope.
During signal analysis, the input impedance drops to 100kΩ Ω Ω Ω , , , , returning to
1000kΩ Ω Ω Ω during the tracing of the graph.
4.2 - Continuous Mode (con)
Place the selector switch in position (osc). If necessary, set the mode (con)
menu of the oscilloscope. Also set the time base, amplitude, trigger type and
trigger level desired (see "Menu Navigation").
After configuration, the signal is captured and continuously updated on the
screen. Touch the probe Ziptec the wire or terminal to be measured and to keep
in touch.
105