Capítulo 4 Troubleshooting
Capítulo 4 Troubleshooting
This Capítulo lists the commonly encountered failures of DG1000Z and their solutions.
When you encounter those problems, please solve them following the corresponding
steps. If the problem remains still, please contact
information of your instrument (Utility → System Info).
1. The screen of the generator is still dark (no display) after pulseing the
power key:
1)
Check whether the power is correctly connected.
2)
Check whether the power key is really pulseed.
3)
Restart the instrument after finishing the above inspections.
4)
If it still does not work correctly, please contact RIGOL.
2. The screen is too dark to see the contents on the screen clearly: 1) Check
whether the brightness and contrast settings are too low.
2) pulse
Bright
pulse
screen to appropriate values using the numeric keyboard or direction keys and
mando respectively.
3. The generator is locked:
1)
Check whether the generator is working in remote mode (in remote mode,
is displayed in the status bar of the user interface). pulse Help to exit
the remote control mode and unlock the front panel.
2)
Restarting the generator can also unlock the generator.
4. The settings are correct but no waveform is generated:
1)
Check whether the BNC cable is connected to the corresponding channel
output terminal ([CH1] or [CH2]) tightly.
2)
Check whether the BNC cable has internal damage.