Chapter 3 Remote Control
2. Configure network parameters 1)
DHCP mode:
If the network supports DHCP, the DHCP server in the network assigns
network parameters (IP Address, Subnet Mask, Gateway and DNS) for
the generator automatically.
2)
Auto IP mode:
When the network doesn't support DHCP, the DHCP mode of the
generator is disabled or the generator is connected to the PC directly, the
generator selects Auto IP mode and acquires the IP address from
169.254.0.1 to 169.254.255.254 and subnet mask 255.255.0.0
automatically.
3)
Manual IP mode:
Enable the Manual IP mode and disable the DHCP mode as well as the
Auto IP mode to set the network parameters manually.
If the generator is connected to the PC directly, set the IP Addresses,
Subnet Masks and Gateways for both of the PC and the generator. The
Subnet Masks and Gateways of the generator and PC must be the same
and the IP Addresses of them must be within the same network segment
(for details, refer to the TCP/IP protocal). An example is shown in the
table below.
Network Parameters
IP Address
Subnet Mask
Default Gateway
If your generator is connected to the LAN of your PC, please acquire valid
network parameters (such as IP Address) and configure the network
parameters of the generator according to the description in "To Set LAN
Parameters".
DG1000Z User's Guide
PC
192.16.3.3
255.255.255.0
192.16.3.1
RIGOL
Generator
192.16.3.8
255.255.255.0
192.16.3.1
3-6