Connecting
3.2 Transmitter Wiring
3.2.1

Wiring Temperature Sensor to Transmitter

Wiring Temperature Sensor to the Analog Input Module
Set transmitter and instrumentation power to OFF when inserting or removing the Analog
Input Module or when making connections to TB1, TB2, TB3 and TB4.
1. Disconnect power from the unit to the transmitter..
2. Open the transmitter top cover by releasing the cover latch.
3. Loosen the captive thumbscrew securing the Access Cover and remove Access Cover.
4. Using a flat-blade screwdriver, remove four captive screws securing the I/O board.
20
DANGER
Remove board and set it aside.
Access Cover Screw
Flowmeter
Power Switch
Figure 3-2
Analog Input Module Access
Latch
Access to Analog Input Module
Operating Instructions, 7/2010, A5E03086486-02
FUS1010 IP65 NEMA 4X