Weinmann VENTI motion BiLevel ST WM 24800 Manual De Instrucciones página 37

Aparato de respiración asistida
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Indicator
IPAPmin (leakage)
Technical alarms
Indicator
Pressure measur-
ing hose discon-
nected!
Device Error!
Excessive pressure!
Disconnection!
Device Error!
Pressure sensor
implausible
Alarm
Minimum therapy
pressure set by
doctor not
reached.
Low priority
Alarm
Low priority
Medium priority
Pressure measur-
ing hose blocked.
Medium priority
Medium priority
Medium priority
Cause
Filter soiled.
Mask leaking.
Masal mask defective. Replace mask
Implausible settings
Cause
Pressure measuring
hose not properly fit-
ted or forgotten dur-
ing assembly
Pressure sensor
defective
Drops of water in
pressure measuring
hose
Hose system not con-
nected to unit prop-
erly or not connected
at all
Unit being operated
with open mask
(mask not on).
Drops of water in
pressure measuring
hose
Rectification
Clean/change filter
Adjust headgear/head-
band so that the mask
does not leak - replace if
necessary
Have settings checked
by your doctor.
Rectification
Check correct fitting/
presence of pressure
measuring hose
Have unit repaired
Dry pressure measuring
hose as described in 6.2
Check connection of
hose to unit
Put on mask or switch off
unit.
Dry pressure measuring
hose as described in 6.2
Troubleshooting
EN
37

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Ventimotion wm 24800

Tabla de contenido