6.2.16.2 Normalize tracking generator

1. Set reference
1.- Press the Option key.
level
2.- Press F1 (TG).
3.- Press F5 (Ref Value).
4.- Set the reference value using the Arrow
key or Scroll knob
Normalization target level is set at Reference value, regardless of
the TG output level.
Range: - 130 ~ +20dBm
2. Run
1.- Press F3 (Execute Normalization).
normalization
2.- Press F2 (Yes) to confirm, or F1 (No) to
cancel. Normalization runs.
3.- Press F6 (Return) to go back to the
previous menu.
3. Activate
1.- To enable normalization, press F4 (Norm
normalization
Corr On).
2.- Normalization is activated and the TG
icon appears.

6.2.16.3 Check TG installation status

1. Run
1.- Press the System key.
normalization
2.- Press F6 (More).
3.- Press F4 (System Config On). The
configuration appears.
4.- The TG sign shows the installation status,
installed
(unchecked).