Using Airvo 2 - F&P Airvo 2 PT101 Serie Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 33

3. USING AIRVO 2

Last
Disinfection:
#16
1. SWITCH ON UNIT
Plug the unit's power cord into the mains/utility power socket. The
connector at the other end of the power cord should be well secured to
the rear of the unit.
WARNINGS
To avoid electric shock:
• Ensure that the unit is dry before plugging into the mains/utility power socket.
Switch on the unit by pressing the On/Off button.
2. CHECK DISINFECTION STATUS
The unit will show you whether it is safe for use on a new patient.
This AIRVO 2 is safe for use on a new patient.
This AIRVO 2 has not been cleaned and disinfected since last use.
This AIRVO 2 is NOT safe for use on a new patient.
3. WARM-UP
The unit will begin to warm up. You will see numbers showing the
current output dew-point temperature, flow and oxygen values. These
numbers will pulse until they approach their target settings.
This screen is called the "Summary screen".
4. JUNIOR MODE
If the patient will be using an Optiflow Junior nasal cannula (OPT316/
OPT318), you must activate Junior Mode.
Junior Mode limits the target settings to: 34 °C and 2 - 25 L/min, in
increments of 1 L/min.
New target
settings
A – 6
To activate Junior Mode:
Hold the Mode button for 5 seconds.
New target settings
The target settings for dew-point temperature and
flow will be changed automatically. The colorful icons
in the corners of the screen indicate that this unit is in
Junior Mode.
To deactivate Junior Mode, follow the same
procedure: hold the Mode button for 5 seconds.