Vent Device Installation; Vent Valve Installation - Varian Turbo-V 2000HT Manual De Instrucciones

Ocultar thumbs Ver también para Turbo-V 2000HT:
Tabla de contenido

VENT DEVICE INSTALLATION

The vent device mod. 969-9831 allows to avoid
undesired venting of the pump during a temporary
power failure (adjustable time up to 36 min.), and
enables an automatic vent operation.
Unscrew and remove the threaded plug (see fig-
ure below).
Screw the flange mod. 969-9108 on the pump,
taking care of the o-ring right position.
Assemble the seal ring and lock the vent device in
position using the KF klamp.

VENT VALVE INSTALLATION

The vent valve mod. 969-9843 allows to avoid un-
desired venting of the pump during a temporary
power failure (5 sec maximum), and enables an
automatic vent operation.
Unscrew the threaded plug (see figure below).
Screw the vent valve into the pump without insert-
ing the supplied adapter between the valve and
the pump.
Tighten the valve using a 16 mm. hexagonal
spanner with a torque of 2.5 Nm.
Do not overtighten the valve as this may damage
the thread on the pump.
30
TECHNICAL INFORMATION
!
CAUTION
87-900-914-01(F)
Tabla de contenido
loading

Este manual también es adecuado para:

969-9084 serie969-9085 serie

Tabla de contenido