3
CAUTION
1. Before starting any maintenance operations, the analyser should be switched off and
disconnected from the power supply by unplugging the power cable. Refer to the
Trouble-Shooting guide (section 6) if the MININEPH
impossible to trace the fault, contact the supplier.
2. Avoid spillage of liquids over the case. Contamination by fluids can damage the optical
system and create a shock hazard. If fluid is spilt over the analyser or into the cuvette
chamber, switch off and wipe up excess fluid. Allow the analyser to dry completely before
switching back on.
3.
Warning: While the MININEPH
samples that are used and all parts and consumables in contact with the samples
must be considered biohazardous. Appropriate precautions must be taken when
working with biohazards. Always wear protective gloves when handling potentially
infectious substances. Technicians must be trained in the safe handling and clean
up of potential blood borne pathogens. Universal precautions, appropriate
hygiene, and decontamination of surfaces are recommended. Consult the reagent
kit insert for precautions on handling potentially hazardous substances.
4. This equipment is earthed and insulated to ensure operator safety. Before use, always
check the condition of the power cable and the plug for wear or damage. The
MININEPH
PLUS
5. This MININEPH
fuses.
6. This product should only be used by suitably trained persons for the purposes stated.
Adherence to the given procedure is recommended. Standard laboratory practice should be
used to ensure safety whilst handling potentially infectious agents.
7. Power supply voltage fluctuations should not exceed +/- 10% of the normal voltage.
8. Transient over-voltages should not exceed 1500V. Using the equipment at other voltages
may cause damage or failure.
9. The precision of the assays performed will be dependent upon the use of well maintained,
calibrated pipettes.
10. Binding Site MININEPH cuvettes are designed for single use only. Re-use will lead to
false results.
11. If this MININEPH
instructions, electrical protection may be impaired.
Insert Code: EIN023
should only be used with the supplied power cable.
is double pole fused. If the MININEPH
PLUS
is not used in accordance with the operating conditions and
PLUS
PLUS
alone does not present a biohazard, the
PLUS
Page 5 of 121
does not function correctly. If it is
fails to operate check both
PLUS
Version: 11
th
August 2011