APPENDIX: 2
Below are guidelines for assays using the User Prepared Chemistry mode of the
MININEPH
PLUS
1.
Use only MININEPH cuvettes.
2.
All sample dilutions must be made using MININEPH Sample Diluent (SN042).
3.
Ensure that the analyser is primed with an appropriate on-line buffer. UPC mode
always draws its on-line buffer from line „B‟.
4.
Begin with a blank time of 5 seconds and a read time of 600 seconds.
5.
The pipette aspiration sequence depends on the chosen assay temperature. For a
low temperature assay, the sequence is:
Aspirate 400L (e.g. Reaction buffer)
Aspirate 40L (e.g. Air gap)
Aspirate 40L (e.g. Reagent)
Dispense and blow out all reagents
For a high temperature assay, the sequence is
Aspirate 40L (e.g. Reagent)
Aspirate 40L (e.g. Air gap)
Aspirate 100L (e.g. Supplementary reagent)
Dispense 480L (All aspirated reagents plus on-board buffer)
6.
Perform assays at doubling dilutions of the assay calibration fluid to determine the
range. Begin with a sample volume of 40L.
When the test range has been determined, select an appropriate sample test dilution
for this range.
If a suitable assay range cannot be determined the following parameters may be
adjusted:
(1) Antiserum dilution
(2) Sample dilution or volume
(3) Assay time (blank time or read time).
7.
The scatter units for the top point of the curve should be in the range 2000-3000. The
scatter units for the bottom point should be between 100-500 units.
8.
Avoid curves that may suffer antigen excess problems. If antigen excess is
suspected repeat the assay using a less dilute sample. If this results in an increase in
the scatter units there is no antigen excess problem.
9.
It is preferable to use a curve-fitting program. Errors for such a curve should be within
specified limits (+/- 5% is recommended). If errors fall outside this range, consider
further changes to the assay conditions.
Insert Code: EIN023
UPC ASSAY DEVELOPMENT
.
Page 29 of 121
th
Version: 11
August 2011