4.4.5 OPERATOR MESSAGES
See also section 6 Trouble Shooting
Message
Invalid card
Wrong version -
contact supplier
Blank too high -
run aborted
Reagent expired
Scatter too high -
reassay
Result
>XXXX
Result
<XXXX
Printer error
Cal Neph
Insert Code: EIN023
Possible
!
Cause(s)
1. Calibration card swiped
incorrectly
2. Calibration card is for a
different assay to that
selected
1. Calibration card swiped
incorrectly
2. Different software versions
on card and analyser
1.Very high analyte
concentration
2. Sample is too turbid for an
accurate determination to be
made
Reagent lot has passed its
recommended use-by date.
The total scatter units have
exceeded the maximum
value of 4000
Result obtained is above the
pre-programmed assay range
Result obtained is below the
pre-programmed assay range
The MININEPH
download the result to the
printer
Analyser calibration failure
Page 20 of 121
Suggested
?
Action
Swipe card again (see
sections 4.4 and 6)
Check chemistry number on
card matches selected assay
Swipe card again (see
sections 4.4 and 6)
Contact your local
MININEPH
Repeat the assay using a
more dilute sample
Check sample is not
lipaemic, icteric or
haemolytic.
Repeat the assay with new
reagent
Repeat the assay using a
more dilute sample
Repeat the assay using a
more dilute sample
Repeat the assay using a
less dilute sample
cannot
Check printer is connected to
PLUS
the MININEPH
green LED is on
DO NOT USE
MININEPH
your local supplier
supplier
PLUS
and the
PLUS
. Contact
PLUS
th
Version: 11
August 2011