Intended Use; Description; Specifications - Binding Site Minineph Plus Manual De Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

1

INTENDED USE

For in vitro diagnostic use
MININEPH
PLUS
using the principle of nephelometry.
2

DESCRIPTION

MININEPH
PLUS
by an antibody/antigen reaction to determine protein concentrations.
The light source is a diode laser that emits at 670nm. The focused light passes through a
cuvette containing the reaction mixture, where antibody/antigen complexes cause light to
be scattered. This scatter is proportional to the amount of antibody/antigen complexes that
have formed, and is detected by a photodiode.
For each assay, a scatter reading is taken at the beginning of the antibody/antigen
reaction (blank or 1st reading), followed by a second scatter reading at a fixed time. The
analyte concentration is calculated using the difference between these two readings.

2.1 SPECIFICATIONS

MEASUREMENT METHOD: Endpoint nephelometry.
LIGHT SOURCE: Laser diode (670nm, <1mW-Class II).
AMBIENT OPERATING TEMPERATURE: 18 - 30°C.
INPUT: Pre-programmed magnetic swipe cards and 15 key membrane keypad
OUTPUT: 2 x 16 character alphanumeric LCD display
INTERFACE: 1 printer interface.
DIMENSIONS:
Analyser
Pipette
POWER REQUIREMENTS: 100 - 240V AC, 50/60Hz, 80VA.
CUVETTES: Binding Site MININEPH cuvettes.
PIPETTE TIPS: Standard 0.2 - 10µL tips
INSTALLATION CATEGORY II (as defined by IEC1010-1 1990).
Insert Code: EIN023
is intended for use in determining protein concentrations in body fluids
is an endpoint nephelometer that uses the measurement of light scattered
Width
240mm
45mm
Depth
Height
280mm
205mm
40mm
240mm
Page 1 of 121
Weight
5.5Kg
0.2Kg
th
Version: 11
August 2011