2.4 Storage Conditions
Inadequate storage conditions may shorten the shelf life and may lead to malfunction of the
product.
• Store at temperatures between -5°C/35°C (23°F/95°F).
• Use the product at room temperature.
• Protect from moisture.
• Store at relative humidity range <75% (non-condensing).
3. STEP-BY-STEP INSTRUCTIONS
SmartLite Pro curing light– Operation at a glance
ON/OFF key
1.1
• "Cure" and "PolyCure" tip
• "Illuminate" (Transillumination) Tip
manner from"low –anterior" to "high-posterior" to OFF mode.
Indicator lights
Light under ON/OFF Key
• Slowly flashing orange SmartLite Pro battery has low power and must be exchanged soon
• Fast flashing orange SmartLite Pro battery is empty and must be exchanged to continue
to operate
• Solid orange SmartLite Pro unit is in an overheating protection state and can't be operated
until light turns off.
Light adjacent charging port
• Solid orange SmartLite Pro curing light battery is charging
• Solid green SmartLite Pro curing light battery is fully charged
Light adjacent radiometer
• Solid red SmartLite Pro curing light output is below 1000 mW/cm
(e.g. wrong positioning
• Solid green indicates irradiance of at least 1000 mW/cm
Audible signals
• One short beep: battery or probe tip insertion into handpiece.
• One beep:
Transillumination Tip: start of cycle, change of power level, interrupt or end of cycle
Cure/PolyCure Tip: start of cycle, interrupt or end of cycle
• Two beeps: warning (i.e. no tip attached)
• Four beeps: overheating protection
Vibratory signals
Transillumination Tip: no vibratory signals
• One vibration:
Cure/PolyCure Tip: start of cycle, interruption of cycle, end of cycle
• Two vibrations:
Cure/PolyCure Tip: warning (i.e. no tip attached)
• Four vibrations:
Cure/PolyCure Tip: overheating protection
10
2
: Starts or disrupts the 10-second curing cycle.
4
: Pressing the ON/OFF key will change in a rotating
1.1
5.5
(
NOTE: no light with battery on top indicates missing contact
5.3
F1
, contaminated, or scratched lens)
and not adequate
2
2
)