2.4 Storage Conditions
Inadequate storage conditions may shorten the shelf life and may lead to malfunction of the
product .
• Store at temperatures between -5°C/35°C (23°F/95°F) .
• Use the product at room temperature .
• Protect from moisture .
• Store at relative humidity range <75% (non-condensing) .
3. STEP-BY-STEP INSTRUCTIONS
SmartLite Pro curing light– Operation at a glance
ON/OFF key
1.1
• "Cure" and "PolyCure" tip
2
: Starts or disrupts the 10-second curing cycle .
• "Illuminate" (Transillumination) Tip
4
: Pressing the ON/OFF key will change in a rotating
manner from"low –anterior" to "high-posterior" to OFF mode .
Indicator lights
Light under ON/OFF Key
1.1
• Slowly flashing orange SmartLite Pro battery has low power and must be exchanged soon
• Fast flashing orange SmartLite Pro battery is empty and must be exchanged to continue
to operate
• Solid orange SmartLite Pro unit is in an overheating protection state and can't be operated
until light turns off .
Light adjacent charging port
5.5
(
NOTE: no light with battery on top indicates missing contact
• Solid orange SmartLite Pro curing light battery is charging
• Solid green SmartLite Pro curing light battery is fully charged
Light adjacent radiometer
5.3
• Solid red SmartLite Pro curing light output is below 1000 mW/cm
(e . g . wrong positioning
F1
, contaminated, or scratched lens)
• Solid green indicates irradiance of at least 1000 mW/cm
Audible signals
• One short beep: battery or probe tip insertion into handpiece .
• One beep:
Transillumination Tip: start of cycle, change of power level, interrupt or end of cycle
Cure/PolyCure Tip: start of cycle, interrupt or end of cycle
• Two beeps: warning (i . e . no tip attached)
• Four beeps: overheating protection
Vibratory signals
Transillumination Tip: no vibratory signals
• One vibration:
Cure/PolyCure Tip: start of cycle, interruption of cycle, end of cycle
• Two vibrations:
Cure/PolyCure Tip: warning (i . e . no tip attached)
• Four vibrations:
Cure/PolyCure Tip: overheating protection
10
)
and not adequate
2
2
Transillumination Tip
Signal Condition
Vibration
Beep
Battery Insertion
-
1x
Start Cycle
-
1x
Second power level
-
1x
Stop cycle manually
-
1x
End of cycle
-
1x
Warnings
-
2x
(i . e . no head attached)
Low battery
-
-
Empty battery
-
-
Overheating protection
-
4x
3.1 Installation and charging
A&B
1 . Insert battery pack into SmartLite Pro handpiece
charged however, it may be necessary to charge the battery before first use .
2 . To recharge the battery:
• Connect the charging base to the Power Cord using the USB mini connector
• Make sure power socket used for power connector is accessible at all times in case of
necessary emergency disconnection .
• Insert the battery onto the charging base
orange, indicating battery is recharging and will show constant green when fully charged .
Note, the SmartLite Pro comes with two batteries . It is recommended that the battery not
being used is stored on the charging base so that it is fully charged when needed .
3.2 Operation – Curing
1 . Select the appropriate LED curing tip for the material to be cured . Attach the tip to the
handpiece by firmly pressing tip into handpiece opening while slightly rotating at the same
moment .
Insufficient Curing – compromised restoration
• Always select the curing tip according to the wavelength of the initiator(s) in the
material to be cured.
• Observe recommended curing time for the chosen tip and curing application.
2 . Protect LED tip from gross debris by applying an FDA-cleared, single use SmartLite Pro
barrier protection sleeve
D
. Make sure that the lens is not blocked by crinkles or seam of
the sleeve
D4
.
Cross-contamination
• Ensure the disposable polyethylene FDA-cleared barrier protection sleeve has
been correctly applied over the entirety of the LED tip and handpiece before
beginning a procedure
D3
.
• The Polyethylene barrier Sleeve does not replace cleaning and disinfection of
the dental instruments. Please clean and reprocess dental instruments after each
patient as described in Section 4, Hygiene and Maintenance
3 . Attach the SmartLite Pro eye protection shields supplied with the unit in combination with
SmartLite Pro barrier protection sleeves . Hold the tip over the opening of the shield and
turn the tip against the shield by 90° into its final position
Pro eye protections shields are securely fixed to SmartLite Pro to avoid accidental aspiration
(press SmartLite Pro eye protection shield firmly into correct position) . Always make sure
SmartLite Pro eye protection shields are properly mounted on SmartLite Pro without
covering the light aperture
E3
.
4 . Use suitable, light filtering safety glasses
Cure / PolyCure Tip
Signal LED
Vibration
Beep
Signal LED
-
-
1x
-
-
1x
1x
-
-
NA
NA
NA
-
1x
1x
-
-
1x
1x
-
-
2x
2x
-
blinking
-
-
blinking
blinking fast
-
-
blinking fast
continuous
4x
4x
continuous
B3
. SmartLite Pro battery pack is pre-
A2
B2
. The battery light
5.5
will illuminate solid
E2
. Always make sure SmartLite
11