6. To utilize a different application during the same treatment visit, the probe tip must be
changed. Remove barrier sleeve. Use moderate force to pull the probe tip from the
handpiece. Place the desired probe tip on the handpiece and press
into place while slightly rotating at the same moment.
barrier if damage is suspected. Attach the light shield when using one of the curing light tips.
7. Clean, disinfect and prepare contaminated handpiece and used tip(s) for reuse according
to section 4, Hygiene and Maintenance.

4. HYGIENE

Cross-contamination.
Infection.
• Do not reuse single use products. Dispose of in accordance with local regulations.
• The barrier is designed for single use and must be disposed of after each use in
accordance with local regulations. The barrier is not a replacement for cleaning,
disinfection, and sterilization.
• Reprocess reusable products as described below.
4.1 SmartLite Pro handpiece
Electrical short-circuit or dangerous malfunction.
Injury.
1. Safeguard handpiece against liquid penetration during cleaning and disinfection.
NOTICE: Wrong cleaning or disinfection method.
Damage to SmartLite Pro curing light.
Instructions for Cleaning and Disinfecting SmartLite Pro Light Handpiece, Tips, Charging Base
Warnings
• The SmartLite Pro Shield should be removed and cleaned / disinfected / sterilized as
outlined below. The SmartLite Pro Light handpiece, tips and charging base are not
sterilizable by autoclave.
• The SmartLite Pro curing light cannot tolerate high-level disinfection procedures.
Intermediate-level disinfection is appropriate for the handpiece, tips and charging base.
• Do not autoclave in steam autoclave.
• Do not clean/disinfect in automated washer/disinfector.
• Do not immerse in liquid.
• Do not clean or disinfect with chlorine bleach/sodium hypochlorite (corrosion of
contacts) or Lysol® Brand I.C. ™ Disinfectant Spray (cracking of charging base).
• Disconnect the power supply plug from the power outlet and charging base unit prior to
cleaning/disinfection.
Limitations on
• Repeated reprocessing has minimum effect on these instruments. End of life is
reprocessing
normally determined by wear and damage due to use.
• Cold liquid immersion disinfection/sterilization, chemical vapor sterilization, and dry
heat sterilization methods have not been tested or validated for efficacy and are not
recommended for use.
Initial treatment at
• Remove SmartLite Pro eye protection Shield. Reprocess as outlined below.
the point of use
• Remove protective barrier sleeve and discard according to local regulations.
• Use a new, clean pair of examination gloves.
• Do not disassemble tip from handpiece at point of use.
• Wipe vigorously with disposable cloth / paper wipe in combination with an alcohol-
based, tuberculocidal, quaternary ammonium solution with a label claim for cleaning
e.g., VoloWipes® Disinfecting/Cleaning/Deodorizing Wipes
• Remove all visible soil, ensuring fluid penetrates all crevices. Use fresh wipes to rub
fluid into the crevices. Do not allow solution to penetrate the casing. Discard used
wipes. Additional wipes may be used.
• Do not remove battery pack from light handpiece. Do not attempt disassembly of
charging base.
• It is recommended that the device be reprocessed as soon as is reasonably practical
following use. • Start reprocessing within 1 hour after use.
• Charging base should be reprocessed as soon as reasonably practical after being
exposed to spatter or spray of body fluids, or touched by contaminated hands or
contaminated light handpiece.
C2