Storz AUTOCON III 400 Manual De Instrucciones página 129

Tabla de contenido
123
Technische Daten
Mode symbol
Designation
Standard AUTO
Bipolar scissors
Micro
Forced
Technical data
CCS ARC
HF voltage form
Control
Bipolar coagulation modes
sinusoidal constant
sinusoidal constant
sinusoidal constant
sinusoidal modulated
1
These max. values are not necessarily
created at rated load.
The HF power limitation is subject to a
tolerance of ± 20%.
Datos técnicos
HF power limitation
Peak
voltage
Effect
Power range
-
5 W – 120 W
150 Vp
-
1 W – 120 W
200 Vp
-
0.1 W – 40 W
90 Vp
-
1 W – 100 W
550 Vp
Default values
Effect
Max. Watt
---
40
---
40
---
10
---
50
Tabla de contenido
loading

Este manual también es adecuado para:

Uh400Uh400 uUh401Uh401 u

Tabla de contenido