5.4.4 Calibration failed
Calibration failed
The calibration failed.
0.0 %
0.0 %
Calibration failed
The calibration failed.
Cancel
5.4.5 Shift detected
Shift detected
Platform shift detected.
0.0 Nm
0.0 Nm
0.0 %
0.0 %
Shift detected
Platform shift detected.
Cancel
743L500 3D L.A.S.A.R. Posture
Live view
Live view
Calibration failed. Tap to recalibrate...
Calibration failed. Tap to recalibrate...
tion failed. Tap to recalibrate...
Cancel
Recalibrate
Nm
Nm
0.0 %
0.0 %
1
2
Recalibrate
Live view
Platform shift. Tap to recalibrate...
Platform shift. Tap to recalibrate...
Cancel
Recalibrate
0.0 Nm
0.0 Nm
0.0 %
0.0 %
1
2
Recalibrate
Calibration in this section deals with the cameras' abil
ity to detect the position of the force measurement
plate.
Error message text:
: "Calibration failed"
: "The calibration failed"
: "Cancel"
: "Recalibrate"
Possible causes:
•
One or more LEDs are not recognised, are covered
or are outside the outer blue frame
•
The camera stand or force measurement plate was
moved during the calibration
•
Overexposed image (white)
•
Underexposed image (black)
•
Backlighting
Measures:
► Carry out the appropriate measures depending on
the cause (for example align the force measurement
plate and camera stands so that all LEDs are recog
nised within the outer blue frame; restart the mini-
computer in case of incorrect exposure).
Take corresponding steps to prevent backlighting
► 1x tap on "Recalibrate".
Error message text:
: "Shift detected"
: "Platform shift detected"
: "Cancel"
: "Recalibrate"
Possible causes:
•
The camera stand or force measurement plate was
moved
•
All LEDs were pushed out of position
•
Underexposed image (black)
Measures:
► Carry out the appropriate measures depending on
the cause (for example align the force measurement
plate and camera stands so that all LEDs are recog
nised within the outer blue frame; restart the mini-
computer in case of incorrect exposure).
► 1x tap on "Recalibrate".
Use