Technical data
Directives and manufacturer's declaration – electromagnetic interference
The 743L500 3D L.A.S.A.R. Posture is designed for operation in an electromagnetic environment as specified
below. The customer or user of the 743L500 3D L.A.S.A.R. Posture must ensure that the device is operated in
such an environment.
Interference measurements
HF emissions according to CIS
PR11
HF emissions according to CIS
PR11
Emission of harmonics according to
IEC 61000-3-2
Emission of voltage fluctuations/flick
er according to IEC 61000-3-3
Directives and manufacturer's declaration – electromagnetic interference immunity
The 743L500 3D L.A.S.A.R. Posture is intended for operation in the electromagnetic environment specified
below. The customer or user of the 743L500 3D L.A.S.A.R. Posture must ensure that the device is operated in
such an environment.
INTERFERENCE
IMMUNITY TESTS
Electrostatic
discharge
(ESD)
according
IEC 61000-4-2
Electrical
fast
transi
ent/bursts
according
IEC 61000-4-4
Surges
according
IEC 61000-4-5
Voltage drops, short inter
ruptions and fluctuations of
the supply voltage accord
ing to IEC 61000-4-11
104
Compliance
Group 1
Class B
Class D
Complies
IEC 60601 test level
±6 kV contact discharge
to
±8 kV air discharge
± 2 kV for power lines
to
±1 kV for input and output
lines
to
±1 kV voltage
(line-to-line voltage)
±2 kV voltage
(line-to-earth voltage)
< 5 % U
T
(> 95 % drop of U
½ period
40 % U
(60 % drop of U
T
for 5 periods
70 % U
T
(30 % drop of U
25 periods
< 5 % U
T
(> 95 % drop of U
) for 5 s
T
Compliance level
±4 kV contact discharge
±4 kV air discharge
A higher compliance level
is not technically possible
because of the compon
ents for patient safety avail
able on the market.
± 2 kV for power lines
±1 kV for input and output
lines
±1 kV voltage
(line-to-line voltage)
±2 kV voltage
(line-to-earth voltage)
< 5 % U
T
) for
(> 95 % drop of U
T
period
)
40 % U
T
T
(60 % drop of U
5 periods
) for
70 % U
T
T
(30 % drop of U
periods
Electromagnetic
environment
directives
The 743L500 3D L.A.S.A.R. Posture
uses HF energy exclusively for its
internal FUNCTIONING. Therefore
its HF emissions are very low and
interference with neighbouring elec
tronic devices is unlikely.
The 743L500 3D L.A.S.A.R. Posture
is intended for use in all facilities
including residential, and those con
nected directly to the PUBLIC SUP
PLY NETWORK which also supplies
buildings used as residences.
ELECTROMAGNETIC
ENVIRONMENT – Direct
ives
Floors should be wood,
concrete or ceramic tile.
The relative humidity must
be at least 30 %.
The quality of the supply
voltage should be equival
ent to a typical commercial
or hospital supply.
The error message from
the section "Platform not
connected" may occur (see
Page 84). If this happens,
complete
described in that section.
The quality of the supply
voltage should be equival
ent to a typical commercial
or hospital supply.
The quality of the supply
) for ½
voltage should be equival
T
ent to a typical commercial
or hospital supply.
) for
T
) for 25
T
743L500 3D L.A.S.A.R. Posture
-