features not accessible are greyed out.
See table on page 34 of user roles and rights.
To take Security settings into use, go to Settings
→ Measurement flow → Maintenance → Admin
settings → Security settings (see Image 19).
Select Security login ON.
Define access to old results.
Admin password can be changed in Manage
admin account. Default admin password is
QRGOSET.
Set up user and supervisor accounts in User
accounts.
Adjust time after a user will be logged out. This
is different to the hibernate settings in Power
save. If Full standby option in Power save
is in use, the user will be logged out when
Settings
Access to
Security login:
old results:
Admin or
OFF
supervisor
ON
All users
Define security settings.
Cancel
24
PRINCIPLES OF OPERATION
QuikRead go Instrument enters sleep mode.
The user can also log out manually.
In case of a lost or forgotten Admin password, re-
quest a temporary password at softwareupdate.
quikread.com. You will need an instrument serial
number to proceed. The temporary password is
valid for one week and can only be used once to
reset the instrument to factory settings.
LIS settings
From the LIS settings option, the LIS settings for
data transfer can be adjusted. The data are trans-
ferred via a serial or a LAN connection. TCP/IP
settings must be supplied before a LAN connec-
tion can function. Contact your supplier for more
details.
10:32
ADMIN SETTINGS :: SECURITY
Manage admin account
User accounts
Log out timeout:
5 min.
OK
Image 19
Security settings
LIS features
Select the following features OFF or ON.
LIS logging. When set ON, LIS traffic will be
logged.
Identify QC result. When set ON, QC results
will have a special identification in LIS com-
munication.
LIS delay. When set ON, there will be a short
delay between consecutive results sent to LIS.
A delay can be useful in solving congestion is-
sues with older serial-type connections.
Additional ID. When set ON, on top of Patient
ID and Operator ID a third sample ID can be
added, e.g. patient birthdate or Doctor ID.
Additional ID can be taken into use also with
QuikRead go instruments not connected to LIS.
POCT1-A2 features
Select the following features OFF or ON.
Operator login. When set ON in POCT1-A2
features settings, all users need to log in al-
ways before starting to use QuikRead go In-
strument. User information has to be provided
by the POCT1-A2 server.
Patient ID validation. When set ON, Patient IDs
will be checked against a downloaded patient
list before sending results to the LIS.
Patient data shown. When set ON, patient data
will be shown before a measurement starts.
Emergency button shown. When set OFF,
emergency measurement is disabled in login
screen. The emergency measurement can be
taken into use also with QuikRead go instru-
ments not connected to LIS.
loading