Usage
Completely slide the PSP into the barrier en-
i
velope The black (inactive) side of the PSP
must be visible
EN
US
Pull off the adhesive strip and close the barrier
i
envelope tightly by pressing together firmly
1
The light protection cover must be disinfected
i
using a disinfection wipe immediately before it
is positioned inside the patient's mouth
(eg with Dürr FD 350)
Alternatively, a spray disinfectant (eg FD 322,
FD 333) can be used on a soft, lint-free cloth
In the case of image plates Plus size 4, place
i
the bite protector around the light protection
cover with image plate if necessary
26
OBJ_DOKU-0000002448-001-mehrere.indd 26
Recording the X-ray image
Place the PSP in the barrier envelope into the
i
patient's mouth
Doing this, make sure that the active side of
the PSP faces the X-ray tube
2
Set the exposure time and setting values on
i
the X-ray unit (see "82 X-ray unit settings")
Record the X-ray image
i
Result:
The image data must be scanned within 30 min-
utes
Preparing for scanning
Remove the PSP with the barrier envelope
i
from the patient's mouth
NOTICE
Damage to the phosphor storage
plate (PSP) caused by a sharp-edged
holding system
Only use holding systems that do not
i
damage the barrier envelope or the
PSP
Do not use holding systems with
i
sharp edges
Wear hand protection
2
te si d e
O p p o si
s tu b e .
to w a rd
2 1 4 2
8 7
4 7 5
D 4 8
a
CAUTION
Light erases the image data on the
phosphor storage plate (PSP)
Never handle exposed PSPs without
i
either a barrier envelope or a plate
guide
Wear hand protection