Troubleshooting
Fault
Top or bottom bulge in the
X-ray image
EN
US
X-ray image is mirror-
inverted
Round shadow on the X-
ray image
Ghosting or double expo-
sure on X-ray image
X-ray image mirrored in
one corner
36
OBJ_DOKU-0000002448-001-mehrere.indd 36
Probable cause
PSP inserted off-center and at an
angle
PSP exposed on the wrong side
Plus ID PSP (with marker) ex-
posed on the wrong side
PSP exposed twice
PSP not sufficiently erased
PSP kinked during X-ray expo-
sure
Solution
Check the error code on the
i
touch screen
Insert the PSP centrally and
i
straight
Insert the PSP correctly in the
i
barrier envelope
Position the PSP correctly
i
When taking an X-ray, make
i
sure that the active side faces
towards the X-ray tube
Only expose the PSP once
i
Check the erasure unit for
i
proper function
Inform a service technician, if the
i
problem persists
Do not kink the PSP
i