8
Commissioning and first
start-up
NOTICE
Short circuit due to the build-up of
condensation
Do not switch on the unit until it has
i
warmed up to room temperature and
is dry
8.1 Installing and configuring the
unit
The unit supports the following imaging pro-
grams:
– DBSWIN
– VistaEasy (ImageBridge, TWAIN compli-
ant third party dental imaging application
software)
Configuring the network
Network configuration
Various options are available for network config-
uration:
– Automatic configuration via DHCP
– Automatic configuration via Auto-IP for direct
connection of device and computer
– Manual configuration
Configure the network settings of the device
i
using the software or, if applicable, the touch
screen
Check the firewall and release the ports, if ap-
i
plicable
Network protocols and ports
Port
Purpose
45123 UDP,
Device recognition
45124 UDP
and configuration
2006 TCP
Device data
514
UDP
Event protocol data
1)
2005 TCP,
Diagnosis
23 TCP
The port can vary depending on the config-
1)
uration
When the unit is first connected to a
computer, it applies the language and
time settings of the computer
2142100008L41 1708V002
OBJ_DOKU-0000002448-001-mehrere.indd 19
Configuring WLAN on the unit
If the unit is to be operated via WLAN, the con-
nection needs to be configured on the unit
In order to establish a secure WLAN con-
nection, we recommend encrypting the
WLAN network with WPA2
The quality and transmission range of the
WLAN connection can be reduced by
ambient conditions (eg thick walls, other
WLAN devices) Mind the signal strength
when you select a suitable location for
set-up
Prerequisites:
– You need to be logged-in on the unit as
Administrator or Service Technician (Settings
> Access Levels > Administrator/Service
Technician)
Request the WLAN settings from your Net-
i
work Administrator
Tap the following on the touch screen: Set-
i
tings > System Settings > Network
Under Interface, select the WLAN option and
i
confirm with OK
Configure the WLAN
i
Confirm with OK
i
Configuring the unit in DBSWIN
The configuration is done with VistaConfig,
which is automatically installed during the instal-
lation of DBSWIN or VistaEasy
Start > All Programs > Air Techniques > Vis-
i
taEasy > VistaConfig
Ser-
vice
Syslog
Telnet,
SSH
Click
i
This updates the list of connected devices
If the device isn't found, enter the IP address
i
manually and click Register device manually
Activate the connected device in the Regis-
i
tered column
You can also register multiple devices
Installation
EN
US
19